RecallDepth

Orthofix TrueLok External Ring Fixation System, Telescopic Linear Distractor, Threaded in the following sizes: 75mm (Catalog Number: 50-1101), 100mm (Catalog Number: 50-1014), 150mm (Catalog Number: 50-1015) and 200mm (Catalog Number: 50-1016), distributed by Orthofix, Inc., McKinney, TX 75069.

Company
Orthofix, Inc
Recall Initiated
September 28, 2007
Posted
February 26, 2008
Terminated
March 20, 2008
Recall Number
Z-0674-2008
Quantity
69 units.
Firm Location
Mckinney, TX

Reason for Recall

Markings reversed: Graduated markings on telescoping bone distractors were reversed, causing the distractors to compress rather than distract the treatment site when adjusted.

Distribution

Nationwide Distribution - USA states: CA, FL, GA, IL, PA, MI, NY and WI.

Lot / Code Info

Lot numbers:  For Catalog #50-1101 (75mm): 075; For Catalog #50-1014 (100mm): 7241, 076 and 50826; For Catalog #50-1015 (150mm): 2664-01, 7343 and 080; For Catalog #50-1016 (200mm): 145, 7342 and 068.

Root Cause

Process control

Action Taken

Orthofix notified sales agents by phone starting on 09/28/07. Sales agents were to return product for replacements.