RecallDepth

BacT/VIEW C.30a Software update (P/N 514515-1) diagnostic data management system, bioMerieux, Inc., P. O. Box 15969, Durham, NC 27704-0969 USA

This recall has been terminated (originally issued March 22, 2007).

Company
Biomerieux, Inc.
Recall Initiated
December 19, 2006
Posted
March 22, 2007
Terminated
December 8, 2009
Recall Number
Z-0644-2007
Quantity
1233 disks
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Bottle data is not sent to BacT/ALERT 3D instrument.

Distribution

Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands.

Lot / Code Info

All BacT/ALERT 3D instruments (version C.30a only, not the C.30 version of software).

Root Cause

Other

Action Taken

A customer notification advising customers of the issues and the precautions were sent on/about December 20, 2006.

More recalls by Biomerieux, Inc.

View all recalls by this company →