Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.
This recall has been terminated (originally issued February 27, 2007).
- Company
- Stryker Leibinger Usa
- Recall Initiated
- December 22, 2006
- Posted
- February 27, 2007
- Terminated
- June 4, 2008
- Recall Number
- Z-0521-2007
- Quantity
- 26
- Firm Location
- Portage, MI
- Official Source
- View on FDA website ↗
Reason for Recall
Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.
Distribution
Nationwide.
Lot / Code Info
Lots G4C00F72HN, G5H00F72HN, G7T00F72HN, G8W00F72HN and G9700F72HN.
Root Cause
Other
Action Taken
Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.