RecallDepth

Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.

This recall has been terminated (originally issued February 27, 2007).

Company
Stryker Leibinger Usa
Recall Initiated
December 22, 2006
Posted
February 27, 2007
Terminated
June 4, 2008
Recall Number
Z-0521-2007
Quantity
26
Firm Location
Portage, MI
Official Source
View on FDA website ↗

Reason for Recall

Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.

Distribution

Nationwide.

Lot / Code Info

Lots G4C00F72HN, G5H00F72HN, G7T00F72HN, G8W00F72HN and G9700F72HN.

Root Cause

Other

Action Taken

Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.

More recalls by Stryker Leibinger Usa

View all recalls by this company →