C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit may be sold individually or with the processing tray assembly (C1140).
- Company
- Steris Corp
- Recall Initiated
- April 23, 2003
- Posted
- May 15, 2003
- Terminated
- February 5, 2004
- Recall Number
- Z-0829-03
- Quantity
- 138 units
- Firm Location
- Mentor, OH
Reason for Recall
The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process.
Distribution
The product was shipped to end-users located nationwide and worldwide.
Lot / Code Info
All codes.
Root Cause
Other
Action Taken
The firm notified their customers by telephone and letter on 4/23/2003.