Uroview 2600 Uroview System

This recall has been terminated (originally issued June 21, 2003).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
May 12, 2003
Posted
June 21, 2003
Terminated
July 7, 2011
Recall Number
Z-0947-03
Quantity
636 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Uncommanded motion of the urological table is possible should the hand control assembly be exposed to liquids.

Distribution

Nationwide. Foreign distribution to Argentina, Australia, Austria, Canada, Germany, Italy, Japan, Mexico, Russia and Switzerland.

Lot / Code Info

Serial numbers beginning with W2 and W9.

Root Cause

Other

Action Taken

A Product Safety Alert letter was sent to all consignees on 5/12/2003 recommending that all hand controls be removed from the units.