Specimen Gate Lab - MSMS Data Suite, Product Code: 5002-0310. The intended use of the product is data management in prenatal and neonatal screening laboratories designed to allow viewing and management of assay results.

This recall has been terminated (originally issued August 25, 2009).

Company
Perkin Elmer
Recall Initiated
January 15, 2009
Posted
August 25, 2009
Terminated
September 9, 2011
Recall Number
Z-1906-2009
Quantity
13 units
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

If Result Codes are used and if they are manually edited, they are correct unless the laboratory specifically uses the View Assay button in the Specimen History screen to open another assay, and then returns back to the original assay

Distribution

Worldwide Distribution -- United States (AL, AZ, CA, CO, CT, FL, GA, IN, KS, KY, MI, MO, OH, OK, OR, TN and TX) and Canada.

Lot / Code Info

Software versions: 1.2 and 1.3.

Root Cause

Software design

Action Taken

Perkin Elmer began notifying US customers January 15, 2009. Notification included instructions for how to avoid the situation and instructions for how to obtain a correction. The firm requested consignees to send in a Response Form by fax (358-2-2678-357) or by email (TurkuQMresponse@perkinelmer.com). For further information, please contact your closest Specimen Gate support center at specimen.gate.support@perkinelmer.com for North America and specimengatesupportfi@perkinelmer.com for outside of North America.