Lap Instruments/Graspers/Forceps - HOT Model #'s: AUT173016 AUT174301 AUT176643 AUT176643 ETH5DCD ETH5DCS ETHEBF01 ETHPOUCH GYR3000PK GYR3005PK GYR3640 and GYR920005PK Laparoscopic electric instruments are designed for use in minimally invasive procedures and/or open surgical procedures to facilitate coagulation, transaction, resection, mobilization and dissection of tissue.

This recall has been terminated (originally issued January 13, 2009).

Company
Sterilmed Inc
Recall Initiated
August 8, 2008
Posted
January 13, 2009
Terminated
April 26, 2011
Recall Number
Z-0560-2009
Quantity
273
Firm Location
Osseo, MN
Official Source
View on FDA website ↗

Reason for Recall

SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable

Distribution

AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA

Root Cause

Packaging process control

Action Taken

Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.