RecallDepth

CAD SCIENCES LLC 3TP TheraMapTM Software, for oncology treatment planning, assessment and monitoring of the patient response.

This recall has been terminated (originally issued January 27, 2009).

Company
Icad, Inc.
Recall Initiated
December 9, 2008
Posted
January 27, 2009
Terminated
January 28, 2016
Recall Number
Z-0653-2009
Quantity
35
Firm Location
Nashua, NH
Official Source
View on FDA website ↗

Reason for Recall

Software modules not approved for this indication.

Distribution

Nationwide distribution to: CA, FL, HI, KS, NH, NJ. NY, MA, MO, MI ,NC, OH, SC, and WI.

Lot / Code Info

All software versions

Root Cause

PMA

Action Taken

iCAD initiated a voluntary field correction by telephone and letter on 12/5/08 by disabling these features at all affected sites. Since Precision Point and TheraMap are controlled by a software license key, iCAD will disable these features by updating the license at the affected sites.