Pyxis Anesthesia System 3500, Cardinal Health, Clinical Technologies and Services, San Diego, CA

Company
Cardinal Health 303 Inc Dba Alaris Products
Recall Initiated
August 13, 2007
Posted
September 26, 2007
Terminated
March 28, 2012
Recall Number
Z-0555-2007
Quantity
369
Firm Location
San Diego, CA

Reason for Recall

System lock up; mini-drawers -under certain circumstances during use, are not accessable, delaying the user from accessing critical medications.

Distribution

Nationwide to Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Illinois, Indiana, Louisiana, Massachusetts, Missouri, Mississippi, Montana, New Jersey, Ohio, Oklahoma, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Virginia, Vermont, and Wyoming.

Lot / Code Info

All serial numbers of the Pyxis Anesthesia Station, Model 3500

Root Cause

Other

Action Taken

The field action consisted of phone calls and a mailing conducted to all customers for the Pyxis Anesthesia System 3500 beginning on 08/13/2007. The customers will be notified by registered return receipt mail by Recall Notifications dated August 14, 2007. Separate letters will be sent to the Director of Pharmacy, Chief of Anesthesia, Director of Risk Management, and Director of Surgery at each facility. Recommended actions will be provided as a short-term fix and a long-term fix of the software will be issued shortly.