The ETI-Mumps IgG Enzyme Immunoassay Test Kit is sold with assay software version 1.0. that is designed to work exclusively with the ETI-Max 3000 automated platform (version 1.51.1).

This recall has been terminated (originally issued February 23, 2006).

Company
Diasorin Inc.
Recall Initiated
August 8, 2005
Posted
February 23, 2006
Terminated
July 22, 2006
Recall Number
Z-0557-06
Quantity
22
Firm Location
Stillwater, MN
Official Source
View on FDA website ↗

Reason for Recall

Potential for false negative results. A software anomaly may occur when performing the ETI-Mumps IgG assay on the ETI-Max 3000 automated platform. The error occurs when running a combined plate on the ETI-Max 3000, with the ETI-Mumps IgG assay as the second assay on a combined plate with ETI-Measles IgG and/or ETI-VZV IgG.

Distribution

Nationwide. AK, CA, CT, DC, KS, MA, NY, NJ, PA, SD, TN, TX.

Lot / Code Info

ETI-Mumps Assay software version 1.0

Root Cause

Other

Action Taken

Consignees were contacted by telephone and then the firm sent a fax titled 'Diasorin Customer Notification' dated 8/8/05 that documented the information as provided by telephone.