Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.

Company
Diasorin Inc.
Recall Initiated
May 20, 2004
Posted
April 13, 2005
Terminated
July 22, 2006
Recall Number
Z-0703-05
Quantity
9
Firm Location
Stillwater, MN

Reason for Recall

When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added.

Distribution

CA, FL, NH, PA, SD, TX, WA

Lot / Code Info

version 7

Root Cause

Other

Action Taken

Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program.