DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164. DiaSorin Stillwater, Minnesota 55082-0285, U.S.A.

Company
Diasorin Inc.
Recall Initiated
May 25, 2004
Posted
April 13, 2005
Terminated
July 22, 2006
Recall Number
Z-0704-05
Quantity
70 kits
Firm Location
Stillwater, MN

Reason for Recall

Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used.

Distribution

Massachusetts, Saudi Arabia, Sweden, Germany, Spain, Belgium, Italy, Camada. Australia, Portugal, Hungary, Israel,

Lot / Code Info

catalog # CA 1725, lot # 04164

Root Cause

Other

Action Taken

Customers were sent notification letters informing them of the product and problem and recommended the destruction of any remaining inventory. Letter also requested the completion and return of the lower portion of the letter.