Zimmer/CAS Power Cord, Sesamoid Plasty, NA, Zimmer/CAS, Montreal (Quebec), Canada; REF 20-8000-070-12. The device is the power cord component for the Sesamoid Plasty CAS workstation which connects the workstation to the power mains.
This recall has been terminated (originally issued December 22, 2009).
- Company
- Zimmer Inc.
- Recall Initiated
- November 2, 2009
- Posted
- December 22, 2009
- Terminated
- October 15, 2010
- Recall Number
- Z-0558-2010
- Quantity
- 71
- Firm Location
- Warsaw, IN
- Official Source
- View on FDA website ↗
Reason for Recall
The power cord female receptacle is not recessed sufficiently and may present a shock hazard.
Distribution
Nationwide.
Lot / Code Info
Workstation serial numbers SP014, SP015, SP019 through SP022, SP025 through SP033, SP035, SP038, SP039, SP043, SP044, SP047, SP049, SP052, SP057, SP058, SP060, SP061, SP062, SP064, SP065, SP069, SP072, SP074 through SP083, SP089, SP090, SP091, SP093, SP094, SP098, SP099, SP121, SP124, SP125, SP127, SP128, SP131 through SP135, SP151 through SP154, SP157, SP175, SP176, SP183, SP184, SP185, SP190 and SP198.
Root Cause
Device Design
Action Taken
Zimmer sales staff were notified by letter dated 11/2/09 and instructed to locate the units and to upgrade the cords and to notify consignees by copy of a letter addressed to risk managers and dated 11/2/09.