Zimmer/CAS Power Cord, Sesamoid Plasty, NA, Zimmer/CAS, Montreal (Quebec), Canada; REF 20-8000-070-12. The device is the power cord component for the Sesamoid Plasty CAS workstation which connects the workstation to the power mains.

This recall has been terminated (originally issued December 22, 2009).

Company
Zimmer Inc.
Recall Initiated
November 2, 2009
Posted
December 22, 2009
Terminated
October 15, 2010
Recall Number
Z-0558-2010
Quantity
71
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The power cord female receptacle is not recessed sufficiently and may present a shock hazard.

Distribution

Nationwide.

Lot / Code Info

Workstation serial numbers SP014, SP015, SP019 through SP022, SP025 through SP033, SP035, SP038, SP039, SP043, SP044, SP047, SP049, SP052, SP057, SP058, SP060, SP061, SP062, SP064, SP065, SP069, SP072, SP074 through SP083, SP089, SP090, SP091, SP093, SP094, SP098, SP099, SP121, SP124, SP125, SP127, SP128, SP131 through SP135, SP151 through SP154, SP157, SP175, SP176, SP183, SP184, SP185, SP190 and SP198.

Root Cause

Device Design

Action Taken

Zimmer sales staff were notified by letter dated 11/2/09 and instructed to locate the units and to upgrade the cords and to notify consignees by copy of a letter addressed to risk managers and dated 11/2/09.