Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories.

This recall has been terminated (originally issued December 2, 2009).

Company
Maquet Inc.
Recall Initiated
October 22, 2009
Posted
December 2, 2009
Terminated
May 18, 2010
Recall Number
Z-0480-2010
Quantity
9 units in US.
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking the accessory to the OR table, being too large.

Distribution

Nationwide Distribution -- USA, including states of SC and PA.

Lot / Code Info

No serial or lot numbers, just a catalogue number.

Root Cause

Device Design

Action Taken

Urgent Device Correction Letters were sent by Federal Express to affected US hospitals on October 22, 2009. The letter stated the technical description, potential hazard, and the recommended precaution. The letter asked customers to immediately verify whether their Connection Bracket is affected thru the FaxBack provided. Questions or need for additional information can be directed to customers' local MAQUET representative.