RecallDepth

Terumo TenderFlow Pediatric Arterial Cannulae, 3.3 mm (10Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813571.

This recall has been terminated (originally issued July 10, 2007).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
May 22, 2007
Posted
July 10, 2007
Terminated
July 7, 2009
Recall Number
Z-0878-2007
Quantity
129
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).

Distribution

Nationwide.

Lot / Code Info

Lot 2120266.

Root Cause

Other

Action Taken

Consignees were notified via phone beginning on 5/22/07 and by follow-up letters sent between 5/23/07 and 6/14/07.

More recalls by Terumo Cardiovascular Systems Corp

View all recalls by this company →