RecallDepth

STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.

Company
Diagnostica Stago, Inc.
Recall Initiated
March 13, 2007
Posted
July 3, 2007
Terminated
January 16, 2009
Recall Number
Z-0853-2007
Quantity
1,098 units
Firm Location
Parsippany, NJ

Reason for Recall

A dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.

Distribution

Hospitals and laboratories Nationwide

Lot / Code Info

Lot -- Exp:  052783 -- April 2007; 061242 -- November 2007; 061981 -- February 2008.

Root Cause

Other

Action Taken

Customers were notified on March 13, 2007 by letter, first class mail.

More recalls by Diagnostica Stago, Inc.