Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
February 1, 2005
Posted
June 30, 2007
Terminated
July 7, 2009
Recall Number
Z-0930-2007
Quantity
416 for both models
Firm Location
Ann Arbor, MI

Reason for Recall

System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.

Distribution

Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.

Lot / Code Info

Serial numbers 0006 through 0066, 0100 through 0144, 0146, 0147 and 0149 may be affected by this issue. Serial number 0046 have been corrected to date.

Root Cause

Other

Action Taken

The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.