Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).
- Recall Initiated
- February 1, 2005
- Posted
- June 30, 2007
- Terminated
- July 7, 2009
- Recall Number
- Z-0930-2007
- Quantity
- 416 for both models
- Firm Location
- Ann Arbor, MI
Reason for Recall
System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.
Distribution
Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.
Lot / Code Info
Serial numbers 0006 through 0066, 0100 through 0144, 0146, 0147 and 0149 may be affected by this issue. Serial number 0046 have been corrected to date.
Root Cause
Other
Action Taken
The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.