Boston Scientific IQ Guide Marker Wire. Catalog No. 38951-01J0, 185 cm, J-Tip, Single Pack. Catalog No. 38951-01J2, 185 cm, J-Tip, 5 pack. Sterile EO. Manufactured for: Boston Scientific Corporation, One Boston Scientific Place, Natick, MN 01760 USA. Boston Scientific Guide Wires are steerable guide wires available in a variety of lengths and diameters. The distal tip is shapeable, or, alternatively, is available in a preshaped ''J'' tip.
This recall has been terminated (originally issued July 12, 2006).
- Company
- Boston Scientific
- Recall Initiated
- May 30, 2006
- Posted
- July 12, 2006
- Terminated
- June 28, 2007
- Recall Number
- Z-1218-06
- Quantity
- 1455 units total
- Firm Location
- Maple Grove, MN
- Official Source
- View on FDA website ↗
Reason for Recall
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
Distribution
Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.
Lot / Code Info
Catalog No. 38951-01J0, Box Lot No. AND Inner-Pouch Lot/Batch #: 8445153/8416208, 8450266/8419531. Catalog No. 38951-01J2, Box Lot No. AND Inner-Pouch Lot/Batch #: 8445220/8416208, 8450269/8419531.
Root Cause
Other
Action Taken
On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.