Stat Profile Critical Care Xpress CCX+ (R) Anaylzer Catalog NUmber: 37555 R= Refurbished
- Company
- Nova Biomedical Corporation
- Recall Initiated
- March 30, 2004
- Posted
- July 20, 2004
- Terminated
- June 17, 2004
- Recall Number
- Z-0829-04
- Quantity
- 7units
- Firm Location
- Waltham, MA
Reason for Recall
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Lot / Code Info
Domestic S/N: Y02902141 Y0290211Z Y02103141 Y02303101 Y02503281 International S/N: Y02603111 Y02603091
Root Cause
Other
Action Taken
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
More recalls by Nova Biomedical Corporation
StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose...
Dec 20, 2024
StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood...
Dec 20, 2024
StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription u...
Dec 20, 2024