RecallDepth

KCI BariAir Therapy System; AC-Powered Adjustable Hospital Bed; Bariatric Bed, manufactured by Kinetic Concepts, Inc., San Antonio, Texas

Company
Kinetic Concepts, Inc
Recall Initiated
October 21, 1998
Posted
January 9, 2003
Terminated
January 28, 2003
Recall Number
Z-0414-03
Quantity
144 Beds (6 of which were out of service prior to correction)
Firm Location
San Antonio, TX

Reason for Recall

Electrical failure of the bed due to fluid egress into the junction box.

Distribution

Nationwide

Lot / Code Info

All serial numbers of beds produced and distributed to the firm''s service centers at the time of correction.

Root Cause

Other

Action Taken

In an October 21, 1998 e-mail the firm requested return of the product for a planned upgrade. The junction box cover was corrected at the same time of the upgrade.

More recalls by Kinetic Concepts, Inc