TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.
- Company
- Kinetic Concepts, Inc
- Recall Initiated
- June 1, 1997
- Posted
- April 10, 2003
- Terminated
- July 8, 2003
- Recall Number
- Z-0718-03
- Quantity
- 4264
- Firm Location
- San Antonio, TX
Reason for Recall
Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge.
Distribution
Nationwide
Lot / Code Info
Serial numbers 0550037 through 0559399.
Root Cause
Other
Action Taken
In approximately November 1997 the firm implemented the TransportAir retrograde for all active units.
More recalls by Kinetic Concepts, Inc
Sterile packaged wound dressing with white and green inner pouch and outer carton labels, labeled in...
May 23, 2006
RotoProne powered patient rotation bed, part #209500, distributed by Kinetic Concepts, Inc., San Ant...
Feb 14, 2006
V.A.C. ATS Canister with Gel (500 mL), Part #M6275063/10.S, distributed by Kinetic Concepts, Inc., S...