Xpert EV Assay, Catalog GXEV-100N-10. The Cepheid Xpert EV assay is a reverse transcription polymerase chain reaction (RT-PCR) using the GeneXpert Dx System for the presumptive qualitative detection of enterovirus (EV) RNA in cerebrospinal fluid (CSF) specimens from individuals with signs and symptoms of meningitis.

Company
Cepheid
Recall Initiated
September 15, 2017
Terminated
July 28, 2023
Recall Number
Z-0028-2018
Quantity
977 kits (9770 units)
Firm Location
Sunnyvale, CA

Reason for Recall

Customers have reported a higher than expected number of invalid test results when using the recalled product per the package insert.

Distribution

USA, Austria, Australia, Belgium, Brazil, Canada, Chile, Colombia, Germany, Spain, France, UK, Ireland, Israel, Italy, South Korea, Lebanon, Netherlands, Philippines, Saudi Arabia, Sweden, Uruguay

Lot / Code Info

batch 1000045409 (lot 06802)

Root Cause

Nonconforming Material/Component

Action Taken

Recall notification letters were sent out to affected consignees on 9/15/17.