RAPTAR Laser Range Finders
- Company
- Potomac River Group, Corp.
- Recall Initiated
- April 21, 2021
- Recall Number
- Z-0001-2022
- Quantity
- 62 units
- Firm Location
- Ashburn, VA
Reason for Recall
RAPTAR units are capable of being field modified to enable the use of the higher power visible and IR laser pointers and the IR Illuminator.
Distribution
US Nationwide
Lot / Code Info
RAPTAR G02 (Wilcox Part #57200G02)
Root Cause
Radiation Control for Health and Safety Act
Action Taken
Potomac River Group (PRG) sent communications to the purchasers of the affected units that were locked in the eye safe mode and distributed without the keys. PRG expects to offer all affected purchasers at least a refund option. PRG s ability to offer a repair option is contingent upon assistance from Wilcox, which is the only entity PRG is aware of that has the ability to render the units compliant with FDA specifications.