OptiVantage DH Power Injection System.

Company
Mallinckrodt Inc
Recall Initiated
November 7, 2006
Posted
December 16, 2006
Terminated
July 21, 2007
Recall Number
Z-0288-2007
Quantity
1348 devices.
Firm Location
Cincinnati, OH

Reason for Recall

The flow rate programmed on the console may be changed without automatically updating the flow rate programmed on the power head. The injector will always inject in accordance with the parameters shown on the power head. If this lack of synchronization occurs and is not noticed by the clinican, contrast would be injected at a higher or lower flow rate than desired.

Distribution

Worldwide distribution ---- including USA and countries of Australia, Belgium, Brazil, Canada, Chile, Columbia, Denmark, Mexico, Panama, Puerto Rico, Singapore and South Africa.

Lot / Code Info

All codes.

Root Cause

Other

Action Taken

The recalling firm sent a letter dated 11/7/2006 to all consignees.