OptiVantage DH Power Injection System.
- Company
- Mallinckrodt Inc
- Recall Initiated
- November 7, 2006
- Posted
- December 16, 2006
- Terminated
- July 21, 2007
- Recall Number
- Z-0288-2007
- Quantity
- 1348 devices.
- Firm Location
- Cincinnati, OH
Reason for Recall
The flow rate programmed on the console may be changed without automatically updating the flow rate programmed on the power head. The injector will always inject in accordance with the parameters shown on the power head. If this lack of synchronization occurs and is not noticed by the clinican, contrast would be injected at a higher or lower flow rate than desired.
Distribution
Worldwide distribution ---- including USA and countries of Australia, Belgium, Brazil, Canada, Chile, Columbia, Denmark, Mexico, Panama, Puerto Rico, Singapore and South Africa.
Lot / Code Info
All codes.
Root Cause
Other
Action Taken
The recalling firm sent a letter dated 11/7/2006 to all consignees.