RecallDepth

The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.

This recall is currently active, issued September 18, 2025. It was issued by Bigtree Sales Inc..

Company
Bigtree Sales Inc.
Recall Initiated
August 29, 2025
Posted
September 18, 2025
Recall Number
Z-2571-2025
Firm Location
Arcadia, CA
Official Source
View on FDA website ↗

Reason for Recall

The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.

Distribution

U.S.A

Lot / Code Info

SKU BT-GERMWAND-LRG-WHT and Testing: September 8-10, 2020

Root Cause

Radiation Control for Health and Safety Act

Action Taken

BigTree Sales, Inc. notified FDA and their customers as required by 21 CFR 1003.10(b), that their electronic product, Portable UV-C Light Sanitizer Wand/Travel UV Sanitizing Wand, model: HH1103/HD-UV-01PL03 (SKU BT-GERMWAND-LRG-WHT), has been discovered to contain a defect related to its safety. Their customer letter was sent with the knowledge of the U.S. Food and Drug Administration (FDA).