Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Right, Trochanteric Nail Component: N/A

Company
Zimmer Inc.
Recall Initiated
December 2, 2025
Posted
December 22, 2025
Recall Number
Z-0972-2026
Quantity
22
Firm Location
Warsaw, IN

Reason for Recall

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Lot / Code Info

Lot Code: Model No 815509420 ; UDI-DI (01)00887868143571(17)340920(10)66717556 ; Lot Number 66717556 Model No 815509420 ; UDI-DI (01)00887868143571(17)340920(10)66884566 ; Lot Number 66884566

Root Cause

Process design

Action Taken

Firm began notifying customers on December 2, 2025 via Urgent Medical Device Recall letters. Distributors are to locate and quarantine affected product in their inventory. They must also return all affected product from their distributorship and from affected hospitals within their territory. Risk Managers are to assist Zimmer Biomet sales representatives in quarantining all affected product. Sales representatives will remove the product from customer facilities. Surgeons are to maintain awareness of the issue, but there are no patient-monitoring instructions related to this recall recommended beyond their existing follow-up schedules.