Kendall SCD Comfort Sleeves Knee Length Small, Pediatric - For use with Kendall SCD SmartFlow, SKU 84021P Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller

Company
Cardinal Health 200, Llc
Recall Initiated
May 13, 2025
Posted
June 6, 2025
Recall Number
Z-1918-2025
Quantity
6517 cases total
Firm Location
Waukegan, IL

Reason for Recall

Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.

Distribution

Domestic US distribution to AK, CA, FL, MA, TX. International distribution to Japan.

Lot / Code Info

UDI-DI (GTIN) 10885380197925 (EA) 50885380197923 (BX) ALL LOTS

Root Cause

Software change control

Action Taken

On May 13, 2025, the firm began to notify customers via Urgent Medical Device Product Correction letters. Customers were informed of the issue, which exclusively impacts use of the Comfort Small Knee Sleeve. Other sleeve sizes are unaffected and operate as intended with the SmartFlow Controller. The product remains operational if the Comfort Small Knee Sleeve and SmartFlow Controller configuration is not utilized during therapy. A software update on the SmartFlow Controller is necessary before using the SmartFlow controller with the Comfort Small Knee sleeve configuration. Customers should discontinue use of the Comfort Small Knee Sleeve with the SmartFlow Controller and utilize a suitable alternative such as SM thigh or foot cuff. Customer service representatives will coordinate with customers to facilitate a software update which will allow the controllers to be utilized with the small knee sleeves without causing the error.

More recalls by Cardinal Health 200, Llc

View all recalls by this company →