Merit Medical, REF: R2007, EsophyX Z+, (1) Fastener Delivery Device, SterileEO, RxOnly

Company
Merit Medical Systems, Inc
Recall Initiated
March 14, 2025
Posted
April 21, 2025
Recall Number
Z-1640-2025
Quantity
1655 units
Firm Location
Jordan, UT

Reason for Recall

Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Egypt, Israel, Saudi Arabia, Spain, Thailand, Turkey, United Arab Emirates.

Lot / Code Info

All Lots/ UDI: 00810275011089

Root Cause

Under Investigation by firm

Action Taken

On March 14, 2025, Merit Medical issued a "Urgent Medical Device Recall Notice via E-Mail, followed by mail to affected consignees. Merit asked consignees to take the following actions: 1. Read and understand this communication and the updated IFU referenced in the attached poster. 2. There is no need to return product to Merit. 3. Post the attached poster on or near the affected products so users are aware of the updated IFU. 4. Discard the current IFU with the product at point of use. 5. Ensure that applicable personnel within your organization are made aware of this communication. 6. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 7. If you require paper IFUs, please contact RESPONSE@merit.com. 8. Please acknowledging you are aware of the information communicated and have completed the actions required of you. 9. If you require additional product training regarding the IFU updates, please contact your Merit Sales Representative to schedule.

More recalls by Merit Medical Systems, Inc

View all recalls by this company →