RecallDepth

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

This recall is currently active, issued November 21, 2024.

Recall Initiated
October 15, 2024
Posted
November 21, 2024
Recall Number
Z-0501-2025
Quantity
20 units
Official Source
View on FDA website ↗

Reason for Recall

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Distribution

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

Lot / Code Info

UDI-DI: 05060462240029 Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537

Root Cause

Device Design

Action Taken

OrganOx issued Important Field Correction Notice letter on 10/15/24 via Fed'X. letter states reason for recall, health risk and action to take: As a result of the investigation into this event, OrganOx has implemented a secondary battery retention mechanism to prevent battery movement due to jarring or impact that could be related to machine transport or movement. The secondary battery retention mechanism will be installed on your machine by OrganOx service personnel no later than 15 December 2024. Recommendations: The metra in your possession continues to be safe and effective in anticipation of this correction and the machine may be used as normal. Should an unexpected rapid decrease in battery voltage occur as noted by alarm from the metra, please determine the impact to the perfusion event and take action if needed. Should this occur, please follow the OrganOx complaints reporting process and quarantine the machine until it can be serviced. Should you have questions about this important Field Correction please contact: Michael Donahue Michael.donahue@organox.com +1 732-948-2175