Volcano Visions PV. 035 Digital IVUS Catheter REF 81234, PN: 300007367341 , Rx Only, Sterile EO, CE 2797

Company
Volcano Corp
Recall Initiated
May 13, 2022
Recall Number
Z-1302-2022
Quantity
7,520 units
Firm Location
San Diego, CA

Reason for Recall

Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications

Distribution

Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.

Lot / Code Info

All Lots containing IFU Part Number 300004856772/A Revision Date: 05/2021 ,; UDI/DI: 00845225002596

Root Cause

Process control

Action Taken

On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions: 1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal. 2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy. 3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred. 4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email: igtdc.r@philips.com

More recalls by Volcano Corp

View all recalls by this company →