Sharesource Connectivity Platform for Use with the Amia Automated PD System

This recall has been terminated (originally issued June 30, 2021).

Company
Baxter Healthcare Corporation
Recall Initiated
June 30, 2021
Terminated
April 10, 2024
Recall Number
Z-2173-2021
Quantity
1 unit
Firm Location
Round Lake, IL
Official Source
View on FDA website ↗

Reason for Recall

Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

Distribution

Worldwide distribution - US Nationwide distribution and the country of Japan.

Lot / Code Info

Product Code 5C9400, UDI: 45413765565692, 55413765565699

Root Cause

Under Investigation by firm

Action Taken

The firm notified its consignees by letter on 06/30/2021. The letter explained the issue with the instructions for use, specifically, That the AMIA Sharesource User Guide describes the function of the UF (ultrafiltration) Limit setting as the ending criteria of the cycler's Extra Last Drain Mode. Baxter will be sending a Follow-Up communication once the AMIA Sharesource User Guide has been updated on the portal.