Medical Action INDUSTRIES INC., Disposable Instrument Pack Curved Hemostat, REF 56224
- Company
- Medical Action Industries Inc
- Recall Initiated
- May 11, 2021
- Terminated
- May 10, 2023
- Recall Number
- Z-1946-2021
- Quantity
- 97 boxes of 20 devices
- Firm Location
- Arden, NC
Reason for Recall
There is a potential for flaking metal and/or brown spots on the hemostats sold as individual units or packaged in the kits
Distribution
US Nationwide distribution in the states of OH, CO, CA, IL, FL, VA, TX, MN, GA, WA, MA, MD, NY, LA, AZ, OR.
Lot / Code Info
Batch Numbers: 0000277592, Exp. 10/12/2025; 0000274266, Exp. 07/21/2025; 0000275098, Exp. 08/21/2025; 0000276380, Exp. 09/24/2025
Root Cause
Process control
Action Taken
Medical Action Industries, Inc. initiated the recall by email on 05/11/2021 Distributors were instructed to notify their customers. The firm is requesting those in possession of the recalled articles discard them.