NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

This recall has been terminated (originally issued May 10, 2021).

Company
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
Recall Initiated
May 10, 2021
Terminated
November 10, 2021
Recall Number
Z-2099-2021
Quantity
71 kits US
Official Source
View on FDA website ↗

Reason for Recall

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples

Distribution

US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX

Lot / Code Info

Batch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066

Root Cause

Under Investigation by firm

Action Taken

Randox notified the distribution center within the USA via email on 5/10/21.The distributor will then contact the customer directly. Letter states reason for recall, health risk, and action to take: Action to be taken: " Please discard previous value sheets and ensure standard value is updated. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director