Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.

Company
R & D Systems, Inc.
Recall Initiated
January 20, 2021
Terminated
February 13, 2023
Recall Number
Z-1255-2021
Quantity
156 kits
Firm Location
Minneapolis, MN

Reason for Recall

R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.

Distribution

USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,

Lot / Code Info

Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604

Root Cause

Under Investigation by firm

Action Taken

Users of these kits are requested to stop using these kits for measuring Human sTfR for diagnostic purposes in patient samples and to discard all unused kits or kit components. Customers were instructed to complete and return the MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form to R&D Systems by February 20, 2021, by emailing the form to techserv@bio-techne.com. For questions regarding this recall, or need to obtain a product credit, contact Bio-Technes Technical Service Department at 1-612-379-2956 or techserv@bio-techne.com.