COAPTITE Injectable Implant, injectable, sterile, non-pyrogenic implant composed of spherical particles (75-125 microns in diameter of calcium hydroxylapatite (CaHA), suspended in aqueous based gel carrier. Gel carrier is composed of sodium carboxymethylcellulose, sterile water for injection, and glycerin. Product is packaged in a 1.0cc syringe and packed in a sterile pouch.

Company
Merz North America, Inc.
Recall Initiated
September 30, 2020
Recall Number
Z-0401-2021
Quantity
2051 units
Firm Location
Franksville, WI

Reason for Recall

The Instructions for Use (IFU) updated 03/02/2020 does not match data listed in the post-approval study.

Distribution

US only distribution - AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV

Lot / Code Info

Model M0068903000, Lots: 100127846, 100127847, 100128709, 100128710, 100129357, 100129614, and 100129615

Root Cause

Labeling design

Action Taken

Merz NA, in cooperation with Boston Scientific Corporation (BSC) have prepared field action communications including a Medical Device Field Notification Letter and Urgent Medical Device Field Notification Response Form. These letters will be sent via FedEx with delivery confirmation. The units at consignees will not be returned but will be furnished with an updated, revision-controlled IFU.