HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium for the qualitative direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA). Not intended to diagnose infection of guide therapy. Further testing required for susceptibility or typing. Container Type: 15 x 100mm monoplate Packaged: 10 plates/sleeve STORAGE: 2-8 degree C on receipt, LIGHT SENSITIVE IVD UDI: (01)08165760201400 (10)447694 (17)191206

Company
Hardy Diagnostics
Recall Initiated
November 14, 2019
Posted
June 9, 2020
Terminated
April 28, 2021
Recall Number
Z-2298-2020
Quantity
3,580 ea (358 kits, pack of 10 in each)
Firm Location
Santa Maria, CA

Reason for Recall

False Positive results due to the MSSA (methicillin-susceptible) strain Staphylococcus aureus (ATCC 29213) breakthrough when this strain should have been inhibited.

Distribution

U.S. Distribution in the states of WI, CA, WA, IA, MT, UT, AZ, WY, KS, MO, OR, TX, VI, including PR and Guam. OUS: British Virgin Islands, Kenya,

Lot / Code Info

Catalogue no.: G307, Lot number 447694 HDx, UDI # 00816576020140

Root Cause

Nonconforming Material/Component

Action Taken

On 11/14/2019, the firm sent a notification of recall letter to customers, informing them of the performance issue of the device lot of the breakthrough on firm's differential chromogenic medium. The customers were advised to discard the remaining stock, and notify the Technical Services Department. If the customer would like replacement or credit, contact Technical Services Department or via e-mail at TechService@HardyDiagnostics.com, or 800-266-2222 option 2 or ext. 5658.