Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device installed inside resident living areas to allow immediate staff response.
- Company
- Securitas Healthcare Llc
- Recall Initiated
- June 20, 2024
- Posted
- July 23, 2024
- Recall Number
- Z-2417-2024
- Quantity
- 1268 units
- Firm Location
- Lincoln, NE
Reason for Recall
There is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery.
Distribution
US: NE, MN, WA, MD, TX, ON, WI, OR, MS, CA, and Canada
Lot / Code Info
All units of the Affected Device Models sold between July 26, 2022 and June 5, 2024
Root Cause
Software Manufacturing/Software Deployment
Action Taken
Securitas Healthcare issued a recall notice on 06/19/2024 to its consignees via USPS certified mail. The notice explained the problem, potential hazard, and requested the following: "Customer Required Actions: 1. Reduce the Missing Device Timeout to 125 Minutes Hours for all Call Stations. Clients should reduce the Missing Device Timeout to 125 minutes so that a Missing Device Alert will be triggered if a Call Station fails to send a supervision message to the Arial Emergency and Nurse Call System within 125 minutes from the last supervision message. Instructions for reducing the Missing Device Timeout to 125 minutes are attached hereto as Appendix 1. 2. Closely Monitor Missing Device Alerts within the Arial Web Application and Immediately Verify Status of Missing Devices and Cease Using any Call Stations That Do Not Reconnect. Clients should closely monitor the "Missing Device Alerts within the Arial Web Application and immediately check the status of Call Stations that trigger a Missing Device Alert and replace the Call Station battery if it is low or fully depleted. Instructions for replacing the Call Station battery are attached hereto as Appendix 2. Instructions for reconnecting missing devices and confirming that the Call Stations are properly transmitting information to the Arial Emergency and Nurse Call System are attached hereto as Appendix 3. Clients should immediately cease using any Call Station that will not reconnect to the Arial Emergency and Nurse Call System. 3. Closely Monitor the Battery Status of all Call Stations in the Arial Web Client and Replace Battery if it is Low . Clients should closely monitor the Call Station Battery Status in the Arial Web Client and immediately replace the battery if it low as described in the documentation. 4. Use Supplementary Communication Devices or Procedures for High Risk Residents. Clients should enact backup procedures, including increased room checks, or providing alternate emergency