Captus 4000e Thyroid Uptake System with Well and 2 inch shielding. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
This recall is currently active, issued May 24, 2024. It was issued by Mirion Technologies (Capintec), Inc..
- Recall Initiated
- April 22, 2024
- Posted
- May 24, 2024
- Recall Number
- Z-1894-2024
- Quantity
- 1,275 units
- Firm Location
- Florham Park, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Complaints of unexpected detachment of the collimator have been reported.
Distribution
Worldwide - US Nationwide distribution and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, FRANCE, GERMANY, GUAM, HONDURAS, HUNGARY, INDIA, ISRAEL, ISREAL, ITALY, JAPAN, KOREA, LATVIA, MALAYSIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, PHILIPPINES , PHILLIPINES, PUERTO RICO, QATAR, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TAIWAN , THAILAND, TURKEY, UNITED KINGDOM, VIETNAM.
Lot / Code Info
Lot Code: Model No: 5430-30154; UDI 0859942006423; S/N range 940001 through 941398.
Root Cause
Device Design
Action Taken
Consignees were sent an URGENT SAFETY NOTICE via email, dated 4/1/24. Consignees are asked to review the provided notice, share the notice with ass affected personnel, instruct personnel to comply with the provided instructions for safe operation and how to perform a retention plate inspection, and to contact Capintec Technical Support if the inspection shows that the collimator is not properly attached. Capintec Technical Support can be reached by phone at 1-800-631-3826 or at 1-201-825-9500, or by email at capintecsupport@mirion.com. Consignees are also asked to confirm receipt of the notification utilizing the provided QR Code or form.