Neodent GM X 6mm Engaging Titanium Base, REF: 9007162

Company
Preat Corp
Recall Initiated
February 21, 2023
Posted
March 24, 2023
Recall Number
Z-1285-2023
Quantity
26
Firm Location
Santa Maria, CA

Reason for Recall

Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the implant, may lead to abutment/screw fracture, which could result in the abutment/screw/crown being ingested by the patient.

Distribution

US Nationwide distribution in the states of TX, OR, UT, NY.

Lot / Code Info

UDI: 00842092161326, Lots: 211891 and 211912

Root Cause

Process design

Action Taken

On 2/21/23, recall notices were mailed to customers who were asked to discard or return affected devices to the recalling firm. In addition, in the event affected product was further distributed, customers were asked to inform their customers, or to provide a customer list so the recalling firm could notify these additional customers. The customer service number was provided: 800-232-7732