ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy

Company
Boston Scientific Corporation
Recall Initiated
March 8, 2022
Recall Number
Z-0915-2022
Quantity
433 5-packs
Firm Location
Maple Grove, MN

Reason for Recall

Pouch seals may be open, compromising sterility.

Distribution

Distribution in the US and territories, Japan, and Canada

Lot / Code Info

ROTAWIRE Drive Floppy, 5 pack, UPN: H74939462005 Individual units within 5-pack are labeled as Single, UPN: H74939462001 Batch numbers: 27622475, 27622478, 27691111, 27698472, 27706785, 27708190, 27756019 27756205, 27756933, and 27756935 GTIN: 08714729996255

Root Cause

Process control

Action Taken

US consignees were mailed an overnight delivery of the customer letter. OUS consignees will be contacted by letter delivery or email. Accounts are asked to return any unused affected devices to the recalling firm. The customer letter provides customers with the product segregation and return instructions and accounts are asked to complete a Reply Verification Tracking Form indicating that they have received the field removal and followed the included instructions.