djosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear Neck Trials with Spring Sides, Lateral, nonsterile, REF S-200636. The firm name on the label is djosurgical, Austin, TX.

Recall Initiated
July 10, 2018
Recall Number
Z-1692-2020
Quantity
28 (14 REF S200635 and 14 S200636)
Firm Location
Austin, TX

Reason for Recall

The retaining ring on the neck trials has a potential to fail interoperatively.

Distribution

Distribution was made to CA, CO, MD, MN, OH, TX, and WA. There was no foreign/military/government distribution.

Lot / Code Info

All Revision C trials. REF S-2000635: 313R1000, 313R1900, 313R1901; REF S-200636: 314R1000, 314R1900, 314R1901.

Root Cause

Process design

Action Taken

The recalling firm issued a letter dated 7/10/2018 to their field representatives (agencies) via email explaining the reason for recall.