KWIK-STIK 2 Pack Trichosporon dermatis derived from ATCC 204094 (Catalog# 0778P, lot 778-65-3)

Company
Microbiologics Inc
Recall Initiated
February 12, 2020
Terminated
March 26, 2021
Recall Number
Z-1690-2020
Quantity
49 units
Firm Location
Saint Cloud, MN

Reason for Recall

KWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis.

Distribution

Nationwide, Including Australia, Brazil, Curacao, Ecuador, Germany, Greece, Italy, Puerto Rico, Singapore, Turkey, United States, Uruguay and Australia.

Lot / Code Info

Catalog Number: 0778P. Lot Number: 778-65-3. UDI 20845357018773.

Root Cause

Nonconforming Material/Component

Action Taken

Microbiologics notified customer on 02/12/2020 via "Urgent Medical Device Correction" letter. The recall letter was emailed to consignees and the consignees were instructed to properly discard any affected products they have. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.