Cautery Tip Cleaner REF: 30500 Product Usage: electrosurgical accessory

This recall has been terminated (originally issued October 16, 2019).

Company
Xodus Medical Inc
Recall Initiated
September 19, 2019
Posted
October 16, 2019
Terminated
March 4, 2024
Recall Number
Z-0151-2020
Quantity
255,000 each
Firm Location
New Kensington, PA
Official Source
View on FDA website ↗

Reason for Recall

The sterile barrier of some devices may have been affected

Distribution

Worldwide Distribution - US Nationwide in the States of : CA, ID, IL, LA, MD, NY, OH, OK, TX, VA, WI and OUS: Malaysia, Kuwait, Australia, Canada, UAE, Ireland, Korea, Saudi Arabia, Japan, Mexico

Lot / Code Info

Lot # 16DEC19 17JAN06 17JAN06A 17JAN23 17FEB27 17MAR13 17MAR13A 17APR03 17APR14 17MAY24 17MAY24A 17MAY24B 17JUN27 17JUL17 17AUG01 17SEP22 17SEP22A 17OCT23 17NOV27 17NOV27A 17DEC15 18JAN19 18JAN19A 18FEB16 18MAR02 18MAR19 18MAR19A ***Updated 11/9/21*** 19MAR14 UDI: Case 20787551000577 and Each 00787551000573

Root Cause

Packaging

Action Taken

Urgent Medical Device Voluntary Recall notification letters dated 9/20/19 were sent to customers. An Urgent Medical Device Voluntary Recall REVISED RECALL NOTICE notification letters were sent to customers on 9/27/19. 1. Immediately check your inventory and quarantine affected stock on hand. 2. The following process shall be undertaken in order to segregate any affected inventory. The existing inventory shall be inspected and any affected product quarantined and destroyed. A. Inspect the Tyvek pouch seal integrity. B. Secondly, the product contained within the pouch may be partially sealed into the Tyvek seal itself. 3. unaffected product shall be distributed with no futher action necessary. 4. For any product that is quarantined and must be destroyed: A. Xodus Medical requires documentation of the affected inventory that is found to be affected and destroyed. Information regarding the following shall be documented and provided to Xodus Medical. B. Product Number; Quantity Affected; Method of Destruction; Evidence of Destruction. It can be photos or videos of product being destroyed 5. Please immediately complete the attached Acknowledgement form even if you do not have any affected stock, and return the acknowledgement to Xodus Medical at the following: Email PLloyd@XodusMedical.com or Fax +1 724-337-1131 For further information and question, please call Paul Lloyd, Vice President, Global QA/RA & Technology, Xodus Medical; +1 724-337-5500; PLloyd@XodusMedical.com Urgent: Medical Device Voluntary Recall REVISED RECALL NOTICE dated 6/8/20 were sent to customers. Xodus Medical has received complaints of some devices where the sterile barrier of some devices may have been affected. Xodus Medical has confirmed instances where the pouch seal was compromised on some of the reported product. In no instance has it been reported to Xodus Medical that a compromise in the sterile barrier has resul