TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.

Company
Coopersurgical, Inc.
Recall Initiated
July 8, 2019
Terminated
May 28, 2020
Recall Number
Z-2364-2019
Quantity
49813
Firm Location
Trumbull, CT

Reason for Recall

The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.

Distribution

Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.

Lot / Code Info

IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390

Root Cause

Error in labeling

Action Taken

On July 8, 2019, the firm began notifying customers of the recall via Urgent Medical Device Correction letters. The letters informed customers that CooperSurgical has updated the IFU to clarify that use of the Transwarmer Mattress with other heat producing devices, such as an incubator, is prohibited. Such use could lead to serious health consequences, such as skin burns. Customers were provided with an updated version of the IFU. Customers were asked to complete and return the acknowledgement and receipt form to the firm. If you have any questions about this recall, please contact CooperSurgical at 203-601-5200 ext. 3300.