RecallDepth

Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L container) The M-30D Diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-3200 and BC-3600 Hematology Analyzer.

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
January 24, 2018
Terminated
April 19, 2019
Recall Number
Z-1320-2018
Quantity
4838
Firm Location
Mahwah, NJ

Reason for Recall

Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet background count.

Distribution

US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .

Lot / Code Info

(1) 105-005405-00 (2 x 5.5L container) Lot Numbers: 2017060906, 2017041207, 2017042707, 2017041104, 2016122301, 2016112901, 2016060401, 2016052701, 2016052101, 2016050301; (2)105-005406-00 (20L container) Lot Numbers: 2017041104, 2016121901, 2016051801, 2016050301

Root Cause

Material/Component Contamination

Action Taken

Letters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.

More recalls by Mindray Ds Usa, Inc. D.B.A. Mindray North America

View all recalls by this company →