Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO

This recall has been terminated (originally issued January 11, 2018).

Company
Trimed Inc.
Recall Initiated
January 11, 2018
Terminated
January 18, 2024
Recall Number
Z-1373-2018
Quantity
1525 devices
Firm Location
Santa Clarita, CA
Official Source
View on FDA website ↗

Reason for Recall

Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.

Distribution

US, China, Latin America

Lot / Code Info

Affected lot numbers: 922361, 922362, 91956, 91957, 92237, 92238, 91858, 92235, 92656, 91856, 91857

Root Cause

Process control

Action Taken

Recall notification letters were sent to affected consignees on January 16, 2018. The consignee letter requests that the consignee return an initial Acknowledgement & Receipt Form to TriMed by January 23, 2018.