EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

Company
Medica Corporation
Recall Initiated
April 11, 2017
Terminated
August 20, 2019
Recall Number
Z-1267-2018
Firm Location
Bedford, MA

Reason for Recall

Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.

Distribution

USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.

Lot / Code Info

Medica s Lot # 16292

Root Cause

Process design

Action Taken

Consignees were notified of the recall via email bulletin dated 4/11/17. The bulletin informs customers of the issues and explains the actions to be taken.