EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

This recall has been terminated (originally issued April 11, 2017).

Company
Medica Corporation
Recall Initiated
April 11, 2017
Terminated
August 20, 2019
Recall Number
Z-1267-2018
Firm Location
Bedford, MA
Official Source
View on FDA website ↗

Reason for Recall

Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.

Distribution

USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.

Lot / Code Info

Medica s Lot # 16292

Root Cause

Process design

Action Taken

Consignees were notified of the recall via email bulletin dated 4/11/17. The bulletin informs customers of the issues and explains the actions to be taken.