Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Company
Elekta, Inc.
Recall Initiated
January 10, 2017
Terminated
August 18, 2020
Recall Number
Z-1044-2017
Quantity
1999
Firm Location
Atlanta, GA

Reason for Recall

Incorrect dose after editing beam number an wedge angle.

Distribution

AZ, AR, CA, CO, CL, GA. IL, IN, LA, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH, OR, PA, SD, TX, UT, VA, WA, WI, Puerto Rico, Algeria, Antiqua & Barbuda, Argentina, Australia, Austria, Bangladesh, Bahrain, Belarus, Botswana, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Laos, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Napal, Namibia, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Peru, Romania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Venezuela, Vietnam, Zimbabwe

Lot / Code Info

Software Builds - V 5.10.00 (including 5.10.01, 5.10.02); V 5.11.00 (including 5.11.01); V 5.20.00

Root Cause

Other

Action Taken

Elekta sent Important Field Safety Notice 382-01-MON-009 to all affected customers on 1/10/2017. The notice informs users of the specific product and version numbers by the issue, and any work arounds that can be used to avoid the issue. The notice informs customers that Elekta Regional Support Staff will contact them directly. The notice also included an acknowledgment form to be returned.

More recalls by Elekta, Inc.

View all recalls by this company →