RecallDepth

BD SurePathTM Preservative Collection Vials, catalog numbers 490505, 490527, 491095, 491096, and 491097. SurePath Preservative Fluid is designed for use with the PrepStain System. SurePath Preservative Fluid is an alcohol-based, preservation solution that serves as a transport, preservative and antibacterial medium for gynecologic specimens.

This recall has been terminated (originally issued February 12, 2016).

Company
Becton Dickinson & Co. 7 Loveton Cir Bd Diagnostic Systems
Recall Initiated
February 12, 2016
Terminated
December 27, 2016
Recall Number
Z-0877-2017
Quantity
16,682.396
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

BD has confirmed that a portion of BD SurePath" collection vials associated with specific production lots may contain caps that are cracked which can lead to leaking of the preservative fluid.

Distribution

Worldwide Distribution - USA (nationwide) Distribution including BB(Barbados) and PR (Puerto Rico) and to the states of : AK, AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and to the countries of: China, Europe, Hong Kong, India, Korea, Mexico, New Zealand, Panama, Singapore, and Taiwan. 480 US customers and 3 US distributors. Product was also shipped to 10 ex-US regions. 15 US government sites received affected product

Lot / Code Info

SurePath Collection Vial Replacement Cap, cat no. 490505 SurePath GYN Preservative Vial Kit, cat no. 490527 SurePath GYN Preservative Vial Kit with Cervex-Brush, cat no. 491095 SurePath GYN Preservative Vial Kit with Cervex-Brush Combi, cat no. 491096 SurePath GYN Preservative Vial Kit with Brush/Spatula, cat no. 491097

Root Cause

Nonconforming Material/Component

Action Taken

BD Life Sciences sent the initial recall notification on February 12, 2016. The distributors were provided with a written communication via email and customers were provided with a written communication via UPS. For further questions, please call (410) 316-4000.

More recalls by Becton Dickinson & Co. 7 Loveton Cir Bd Diagnostic Systems

View all recalls by this company →