RecallDepth

Alinity m System, Part No. 08N53-002

Company
Abbott Molecular, Inc.
Recall Initiated
December 6, 2021
Recall Number
Z-0460-2022
Quantity
742 devices
Firm Location
Des Plaines, IL

Reason for Recall

There is a potential for abnormal (non-sigmoidal) amplification curves.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambique, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Uganda, United Kingdom, and Vietnam.

Lot / Code Info

Model 08N53-002, all serial numbers

Root Cause

Under Investigation by firm

Action Taken

On December 06, 2021 Abbott Molecular issued a Product Correction letter to all customers who received an Alinity m instrument (List Number 08N53 002). Customers are instructed to follow the Necessary Actions outlined in the Customer Letter until an Abbott Molecular representative adjusts the customers clamp bar. Abbott personnel will follow up with customers by e mailing and or follow up visits/phone calls.

More recalls by Abbott Molecular, Inc.

View all recalls by this company →